8.3.3.31 Protecting Power and Ethernet Cables on the Floor

Procedure
The power cable and the Ethernet cable can be
routed in the left-hand side of the cable tray
behind the table console.
At the end facing the wall, the cable tray has a
cutout that will serve as an exit point for the
two cables.
3 m (10 ft) of rubber cable protectors is
provided to cover:
Power and network cables running from
the gantry to the wall
Cable between the gantry and acquisition
PC.
Routing and protecting the acquisition PC-
gantry cable is described above (see page 8-
46, 8-57 and 8-63).
The power and network cables running from
the gantry to the wall must also be protected
where traffic across the cables is to be
expected.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Routing power and
Ethernet cable
Cable exit
CardioMD Service Manual 8-73
Installation