Installation

12.6 Installation
This section provides instructions for installing the CardioMD AC Option. For information on
installing the CardioMD system, refer to Chapter 8 Installation Procedure of the CardioMD
Service Manual. Instructions for operating the CardioMD system, including the AC Option, are
provided in the accompanying CardioMD Operator's Manual.

12.6.1 Preparing for the Installation

This section provides instructions for unpacking the CardioMD AC Option.
The main parts of CardioMD AC Option are:
9ACO1347 Transmission scanner for detector 1
9ACO1371 Transmission scanner for detector 2
A kit for upgrading a CardioMD system to support attenuation correction. This kit includes:
-
Electrical parts and cables for the CardioMD gantry
-
Electrical parts and cables for the CardioMD detector
Bushings with bolts and washers forming the mechanical attachment for the transmission
scanners on the CardioMD detectors
9CSY1381 FOV mask
9CSY0906 CardioMD software installation CD
1MAN0153 CardioMD Operator's Manual and 1MAN0158 CardioMD Service Manual.
Note. Line sources are not included with the CardioMD AC option but are delivered, mounted in
line source housings, directly from the line source vendor.
Note. The CardioMD AC Option requires EDC module firmware revision 7CTL0095-E05 or
later. The firmware version can be determined from the system log file. Check the most recent file
in the directory

12.6.1.1 Shipping Container and Weight

The CardioMD AC option is shipped in a single container. Table 12.6 lists size and weight,
including pallet.
Shipping box with
CardioMD AC Option complete
Table 12.6 Shipping container, weight and dimensions

12.6.1.2 Tools

Tools Shipped with the AC Option
None.
Tools not Shipped with the AC Option
3 and 5 mm standard screwdrivers
Wrench: 5 and 7 mm
Metric Allen keys: 2, 2.5, 3 and 5 mm
Torx key No. 9
A pair of pliers.
12-88 CardioMD Service Manual
C:\cardiocam\SystemLog
.
Shipping weight
122 lbs.
55 kg
×
×
Size W
D
H
31.5 × 15.7 × 25.6 inch
80 × 40 × 65 cm
1MAN0158-H11 04-2005 CSIP Level 1