Repair Procedures

12.5.2.3 Replacing the Motor Power Board

Procedure
1. Unplug the detector and line source
compartment cables.
2. To remove the motor power board,
unscrew the 7 screws from the rear of the
inner bracket. Use a 2 mm Allen key.

12.5.2.4 Replacing Limit Switches

Procedure
1. Unplug the switches from motor power
board.
2. Use a Torx T6 key to unscrew the limit
switches from the motor and gear
assembly.
The photos on the right show the limit
switches on both sides of the inner bracket
assembly.