When the processing station has been configured, also check the DICOM transfer by exporting
some of the test images located in the patient database on the acquisition PC. Follow instructions
provided in the CardioMD Operator's Manual, Chapter 4 Operating Instructions.
DICOM Setup of JETStream®
To configure the JETStream Workspace properly for DICOM transfer from CardioMD, you must
edit the
customMap.ini
Procedure
1. In the Windows Start menu, select
Programs, Accessories, Notepad.
The Windows Notepad application starts.
2. In the Notepad window, select File, Open.
3. Select the file
c:\Program
Files\Philips Nuclear
Medicine\Jetstream
Workspace\dicom\customMap.ini
4. In this file, find the lines:
[Philips_CardioMD]
NameKey = 0018,1030
ExamKey
= 0018,1030
5. Replace them with:
[Philips_CardioMD]
NameKey = 0008,103e
ExamKey = 0018,1030
8.3.4.3

Configuration of Regional Settings

The acquisition PC must be set up to use the US date format (yy-mm-dd) or the international date
format (dd-mm-yy).
Procedure
1. Right-click the Windows Start button and
select Explore from the pop-up menu.
2. In the Explorer's folder list, select the
Control Panel.
3. In the file list (the right-hand pane of the
Explorer window), double-click the icon
Regional and Language Options.
In Windows 2000, the name of the icon is
Regional Options.
1MAN0158-H11 04-2005 CSIP Level 1
file.
Details
.
Details
Installation
CardioMD Service Manual 8-81