Detailed Description
Detector
Table 12.5 Commands for manual positioning of line source housing
Note. In Table 12.5, some of the commands to be sent to the line source motion controller are
printed on a gray background. These commands serve to reset the motion controller and need only
be sent to the motion controller once.
Solenoid Driver
The driver for the solenoid activating the shutter (see Figure 12.18, page 12-20) is a switch-mode
type, powered from the E-Stop controlled 48 V supply. For the first 0.5 second after the driver is
activated, it operates at 100 % duty cycle. After this, it operates at 27 % duty cycle. The solenoid
driver is activated from an output port on the line source motion controller.

12.3.5 CardioMD Detector and Table Console

When the CardioMD AC Option is installed, the CardioMD detector and table console are
provided with additional 3 dedicated circuit boards, a 48 V PSU and associated cables. See AC
installation chapter for details.
12-26 CardioMD Service Manual
Move line source to
Command sequence
400 mm
-> canmsg 533 0x80 0
(position furthest
-> canmsg 533 0x06 0
from motion
-> canmsg 533 0x0F 0
controller)
-> canmsg 533 0x0F 0
-> canmsg 1301 0 0xc0 3 0 0 0xc0 3 0
-> canmsg 1045 0xaa 0xaa 0x10 0 0xaa 0xaa 2 0
-> canmsg 533 0x1f 0
-> canmsg 533 0x6f 0
1MAN0158-H11 04-2005 CSIP Level 1