8.3.3.16 Leveling the Gantry Base

Procedure
1. Using an 8 mm Allen wrench, remove the
4 mounting screws in the gantry frame in
order to get access to the leveling screws.
2. Using a level and the adjusting tool, level
the side bar opposite the table base.
3. Level both end bars, and check the side
bar, where the table base is mounted.
4. Reinsert the 4 mounting screws and
tighten them.

8.3.3.17 Adjusting Center Leveling Screws

Procedure
1. Remove the mounting screw in the middle
of one of the sidebars and adjust the
leveling screw until the plate is just
supported on floor.
2. Remount the mounting screw.
3. Perform the same adjustment on the other
sidebar.
Note. Check this once again. It is very important that the 2 sidebars are in level and parallel
(+/– 0.05°).
1MAN0158-H11 04-2005 CSIP Level 1