Lubrication; Vertical Drive - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Planned Maintenance Procedures
Verify that the rotational motion axis does not enter into any of the ultimate limits.
If an ultimate limit is reached, refer to Chapter 4 Calibration for procedures for calibration of
absolute encoders and ultimate limits.
Also verify that the detectors are automatically stopped before hitting the patient table when driv-
ing the detectors vertically and horizontally towards the patient table.
If it is possible to drive the detectors into the patient table, refer to Chapter 4 Calibration for pro-
cedures to calibrate the absolute position encoders.

11.3.4 Lubrication

11.3.4.1 Tools Required
Grease gun with lithium-based grease.
11.3.4.2 Horizontal Drive
Procedure
1. Remove the cover from the gantry base to
access the horizontal drive.
2. Apply lithium grease with the grease gun
on the zerk fitting shown in the photo to
the right.
3. Also apply a small amount of grease to the
rails inside the gantry base and to the two
gear wheels at the motor.

11.3.4.3 Vertical Drive

Procedure
1. Remove the cover on the side of the Y-
pillar.
11-4 CardioMD Service Manual
Details
Details
1MAN0158-H11 04-2005 CSIP Level 1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents