Procedure
acquisition PC.
6. Open the latest log file in the acquisition
PC directory
C:\cardiocam\SystemLog
7. Check that both motion controllers have
the same software version.
See Chapter 7 Software Update Procedures,
the section Checking Installed Software
Versions, for detailed instructions.
8. If they do not, update motion controller
software.
See Chapter 7 Software Update Procedures
for instructions.
6.2.7

Replacing the Power Supply

Procedure
To replace the AC mains power supply:
1. Unplug all cables to the power supply.
2. Unscrew the four bolts fixating the power
supply to the mounting plate inside the
table console.
Note. The mounting plate works as heat sink
for the power supply. The power supply must
be firmly mounted on the plate when powered.