NEMA Multiple Window Spatial Registration (MWSR)

10.5 NEMA Multiple Window Spatial Registration (MWSR)

10.5.1 Test Requirements

Detector:
Source:
Source holder:
Phantoms:
Total test time:

10.5.2 Test Procedure

Note. If the CardioMD system is equipped with the AC Option, the transmission scanners must be
removed before the NEMA MWSR test can be performed. See Chapter 12 CardioMD AC Option
for instructions.
Procedure
1. In the NEMA menu, select MWSR.
The MWSR page appears.
2. Enter isotope energies as follows:
For Ba enter:
81 in the Energy low field.
155 in the Energy high field.
For Ga enter:
92 in the Energy low field
185 in the Energy high field.
For Tl enter:
72 in the Energy low field
167 in the Energy high field.
3. Click Continue.
4. Click OK to accept automated motions.
Automated motions will now position the de-
tectors for measuring MWSR.
Note. Automated motions must be started from patient load position.
Note. The position used for MWSR measurement is the same for both detectors. In this position
the detectors are raised above the patient table and rotated to +135°.
10-12 CardioMD Service Manual
Phantoms. One detector at a time.
4 sources, Ba-133, Ga-67 or Tl-201 ~ 7–20 MBq or 200–
500 µCi, yielding a total of 20 kcps or less.
4 point holders for the GEOCAL mask.
GEOCAL mask
Approximately 10 minutes per detector.