Gantry Repairs
Open the sleeve and pull the inner conduit with the cables through.
3. Remove the faulty cable from the inner conduit through the sleeve and insert the new cable in
the same manner.
4. Put back the outer conduit and connect the combined conduit at both ends.
6.2.6

Replacing a Motion Controller

Procedure
To replace a motion controller:
1. Unplug all cables to the motion controller
2. Unscrew the two Phillips screws as shown
in the photo on the right.
Take the motion controller out of its chassis:
3. Carefully pull out the end where the
Phillips screws came out by about 3 cm.
4. Slide the other end out of the chassis (two
latches are holding this end down).
When a replacement motion controller is installed, its DIP switches must be set according to its
allocation. See Table 6.1.
Switch #
1
2
3
4
5
6
7
8
9
10
Table 6.1 Motion controller DIP switch settings
Procedure
After replacing the motion controller, check
whether it has the correct software:
5. Power up the CardioMD system and the
6-4 CardioMD Service Manual
Details
Master motion controller
ON (Up)
ON (Up)
OFF (Down)
ON (Up)
ON (Up)
OFF (Down)
OFF (Down)
OFF (Down)
OFF (Down)
OFF (Down)
Details
Slave motion controller
OFF (Down)
OFF (Down)
ON (Up)
ON (Up)
ON (Up)
OFF (Down)
OFF (Down)
OFF (Down)
OFF (Down)
OFF (Down)
Screws to
be removed
1MAN0158-H11 04-2005 CSIP Level 1