Removing Gantry Shipping Pallet Using Dolly Beams - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Procedure
4. To remove the large shipping bracket,
remove 2 bolts and 1 screw at each end of
the bracket.
8.3.3.3

Removing Gantry Shipping Pallet using Dolly Beams

Procedure
1. Locate the 2 jack beams, the 2 wheel bars,
4 plastic protective feet, 4 screws M12 ×
75 mm, and a ½" ratchet.
2. Using 4 pcs. M12 × 75 mm socket head
cap screw and a 10 mm Allen wrench,
mount a jack beam in each end of the
gantry. The side with protective rubber
against the painted surface of the end bar.
3. Apply a plastic protective foot under each
acme screw.
4. Using a ½" ratchet, raise the gantry by
app. 1 inch (25 mm).
1MAN0158-H11 04-2005 CSIP Level 1
Details
Details
Installation
Remove screw
Remove bolts
CardioMD Service Manual 8-27

Hide quick links:

Advertisement

Table of Contents
loading