Internal LED Indicators
5.3.2

LED's on the Detector Power Supply

Five LED's inside the detector power supply provide information about the integrity of the output
voltages.
The LED's are visible through the perforations of the top cover.
Figure 5.6 LED's on the detector power supply
LED
HVHIGH
HVGOOD
HVLOW
LVGOOD
LVLOW
Table 5.1 Significance of LED's on the detector power supply
5.3.3

LED's on the Detector EDC Boards

An LED is located on the rear panel of the EDC board (see Figure 5.7 on page 5-9). The LED of
the EDC board configured as DET0 (detector 1) should be flashing green at 0.5 Hz indicating that
the EDC board is up running. The LED on the EDC board configured as DET1 (detector 2) should
be flashing red at the same rate. At power up or reset, the LED will be constantly red.
A flashing red light indicates that an error was detected during power up.
5-8 CardioMD Service Manual
COLOR
SIGNIFIES
RED
High voltage higher than specified setting
GREEN
High voltage at specified setting
RED
High voltage lower than specified setting or not enabled
GREEN
All low voltages in specified range
RED
One or more low voltages lower than specified range