Gantry
Figure 3.5 Electronics behind front cover
Safety Board
Figure 3.6 Electronics behind rear cover
3.2.1

AC Power Supply

The gantry power supply is located in the lower left corner in the rear of the table console. It
powers the entire system except for the processing station. The outputs are divided into two
groups:
One group of continuous outputs (see Table 3.1 on page 3-7)
One group of safety interrupted outputs (see Table 3.2 on page 3-7).
All low voltage outputs are protected by PTC fuses.
Note. To reset the PTC fuses, all loads on the affected output must be removed; power must be
removed momentarily and then reapplied. Any capacitance within the circuitry will cause the fuse
not to reset. These fuses are not accessible and are not field-replaceable. Should one of these fuses
fail, the entire power supply must be replaced.
3-6 CardioMD Service Manual
Linear Motion
Controller
(Y and X Axes)
Y Drive Motor
Amplifier
100V
120V
200V
220V
12A
12A
6.25A
6.25A 6.25A
FUSE
I
0
FUSE
Rotate Motion
Controller
(Rotate and Table)
Rotate Drive Motor
Amplifier
Secondary fuses
10A both
MAINS VOLTAGE
SETTING
240V
FUSE RATING
( both fuses )
1MAN0158-H11 04-2005 CSIP Level 1