Philips CARDIOMD Service Manual page 227

Table of Contents

Advertisement

Procedure
The table base is shipped in a second wooden
crate (31.5 × 47.2 × 36.6 inch / 80 × 120 × 93
cm).
3. Remove the plywood sides and top.
4. Remove the protection bag by carefully
cutting around the edges.
The acquisition laptop PC, system accessories,
covers and installation Dolly are all shipped
in a third wooden crate (31.5 × 57 ×34 inch /
80 × 145 × 87 cm).
5. Remove the plywood sides and top.
6. Remove the protection bag by carefully
cutting around the edges.
The patient table option is shipped in two
cardboard boxes.
The pallet and mattress are shipped in one
box:
19.7 × 110.6 × 5.9 inch / 50 × 281 × 15 cm
for the long table,
19.7 × 81.5 × 5.9 inch / 50 × 207 × 15 cm
for the short table.
The table end support is shipped in a second
box (31.5 × 23.6 × 31.5 inch / 80 × 60 × 80
cm).
7. Open the boxes and remove the protection
bags by carefully cutting around the edges.
The laptop PC mounting option is shipped in a
single box.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Installation
CardioMD Service Manual 8-25