10.1 Introduction
This section of the CardioMD service manual describes procedures for performing NEMA per-
formance measurements on the CardioMD detectors.
The acquisition PC includes dedicated software to perform these NEMA measurements. To access
any NEMA tests, follow instructions below.
Note. If the CardioMD system is equipped with the AC Option, the transmission scanners must be
removed before NEMA testing can be performed. See Chapter 12 CardioMD AC Option for in-
structions.
Note. The detectors must be uncollimated for all NEMA performance measurements.
Note. Make sure that the system is in the patient load position before accepting any automatic
motions to bring the detectors into position for NEMA measurements.
Procedure
1. If the CardioMD system is equipped with
the AC Option, remove the transmission
scanners before performing NEMA tests.
See Chapter 12 CardioMD AC Option for
instructions.
2. Make sure that the system is in patient
load position before entering any NEMA
procedure.
Automated motion bringing the detectors into
position for the NEMA measurements may fail
if not executed from patient load position.
3. On the Persistence page, click the Tools
button.
4. When the Calibration & Configuration
menu appears, click Service.
5. If prompted, enter the service password
(see Chapter 4 Calibration for instruc-
tions).
6. In the Service Calibrations menu, click
NEMA.
10-2 CardioMD Service Manual
Details
1MAN0158-H11 04-2005 CSIP Level 1