Motion Controller Software Older Than 7Mot0096-N14 - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Procedure
tion controller (motion controller to the
left – SW10 in position down).
Close down the CardioMD application and
then restart it:
6. Ensure that the acquisition PC is display-
ing the Persistence page. Click the Close
button in the upper right corner of the
window.
7. Wait for at least 10 seconds.
8. Start the CardioMD application by double-
clicking the CardioMD icon on the desk-
top.
4.5.4.2

Motion Controller Software older than 7MOT0096-N14

The ranges within which the ultimate switches must go active are:
Horizontal:
Vertical:
Rotational:
Note. After testing ultimate switches, both motion controllers must be reset and the CardioMD
application software restarted in order to ensure correct display of detector rotation.
Procedure
1. Before verifying ultimate switches, set
SW8 upwards on both motion controllers.
It is not necessary to re-boot after changing the
state of SW8.
Procedure
When SW8 is activated, it is possible to use
the hand controller Collision Override button
to position the gantry outside the software
limits and thus reach the ultimate limits. Using
the Motion Enable button as normally makes
the gantry stop at the software limits. Thus, the
position of the limits can be verified one by
one, for example using the sequence outlined
for check of software limits in section 4.5.3.
1MAN0158-H11 04-2005 CSIP Level 1
Details
– 7 mm < X
< – 2 mm, 497 mm < X
min
218 mm < Y
< 228 mm, 1195 mm < Y
min
Z
< –92°, Z
> 182°.
min
max
Details
WARNING
In this mode, all safety/collision software is disabled, so the system must be
operated with extreme care.
Details
< 505 mm
max
< 1199 mm
max
CardioMD Service Manual 4-45
Calibration