2.2 Final assembly

Duo Spray chamber
Figure 13 – Duo Spray chamber assembly
Attach the swan neck fitting to the Spray chamber with the fitting clamp provided. The
swan neck provided with the instrument is specially designed to prevent UV radiation from
escaping from the torch box.
Figure 14 – Duo Spray Chamber fitted
Insert swan neck fitting into torch assembly holder as far as it will go, and connect
up the Nebuliser gas supply to the push-fit fitting.
Warning: It is extremely important that the correct Thermo part is used for the
swan neck. In addition, systems interlocks on the torch holder and elsewhere are
there for safety reasons and must not be bypassed. Operators could be exposed to
dangerous UV and radio frequency radiation if alternate parts are used for the
swan neck.