iCAP Customer Training Manual
Type admin in user name field and click on OK or press enter
4.3.2 Striking the Plasma
Click on the Plasma Icon and then make sure the
interlock indicators all show green, click on Plasma
On. After the plasma strikes, wait until the
Spectrometer Optimization is completed and then
click on the close box
Note: To allow the plasma to stabilise leave Plasma
on with blank solution running for about 10 minutes
before carrying out an analysis.
4.3.3 Setting up Analyses
Click on the Analyst icon
Either select one of the stored methods or create a
New method. To create a new method click on
Cancel
Click on the Method drop down
menu
Click on New and select the
Elements required from the Periodic
table.
To select suitable lines from the list
displayed click OK and OK again