Sample Introduction Glassware Assembly

Warning: Appropriate care and safety procedures should be followed to
avoid
breaking any glassware and causing injury to the operator. Broken
glassware should be handled with appropriate care.

2.1 Torch Assembly

Figure 1 Duo Torch Assembly – Metal Torch Mount
Figure 2 Radial Torch Assembly – Metal Torch Mount
Figure 3 Torch Assembly – torch body
The quartz body of the torch should be pushed fully into the metal torch mount.
The O-rings in the metal torch mount (internally and externally) should be inspected and
replaced if any wear, or damage, is visible.
The quartz body of the torch should be pushed into the metal torch mount. Ensure the
torch body is pushed fully into the metal torch mount. The marked symbol on the quartz
torch is used to locate the radial view hole correctly and should touch the metal body.