Lot Management (Admin Access Only Or Via Poct Server) - Abbott ID NOW NAT-024 Instrument User Manual

Hide thumbs Also See for ID NOW NAT-024:
Table of Contents

Advertisement

QC Lockout - ID NOW Assay 1
Warn
Lockout
Units
Frequency
Cancel
QC Lockout
ID NOW Assay 1
Warn - 100 Tests
ID NOW Assay 2
Lockout - 30 Days
ID NOW Assay 3
O
When the QC Lockout function is enabled (warn or lockout) the user will be
prompted to run QC negative and positive controls before a patient test can be
run. Alternatively, the user can change to a different lot (which has had a valid QC
negative and positive control associated with it.)
Run QC Test
Positive QC Test
Negative QC Test

4.16 Lot Management (Admin Access Only or via POCT server)

Home>Setup>Lot Management
This feature allows the Admin to review detailed Lot information.
Setup
Remote Mode
Connectivity
QC Lockout
Lot Management
The screen will indicate selected Units and
Frequency. Press Cancel to return to previous
screen. When parameters are set as desired, press
O
OK to confirm and continue.
Days
100
OK
The QC Lockout screen will provide a summary for
each assay enabled on the instrument.
Both QC negative and QC positive tests must
generate valid results before the instrument can run
a patient test. Select Run QC Test and follow the
steps in Section 7.
Select Lot Management to review QC test of lots
on instrument as well as to Export and/or Import
lots to and from instrument
28 ID NOW Instrument User Manual