WARNING:
When alcohol purging an endoscope with an elevator, make sure the forceps elevator is in the half-up
half-down position. Follow the endoscope manufacturer's instructions for flushing alcohol into and
around the forceps elevator recess at the distal tip. Endoscope manufacturer may require forceps
elevator deflection during this step.
CAUTION:
The System 83 Plus™ is equipped with a circuit breaker switch that is located on the right hand side if
facing the device. Do not place the breaker in the off position, without first properly shutting down the
computer system to avoid possibly corrupting the system software. The circuit breaker switch is for
system protection, emergency shut down, and service technician use only.
CAUTION:
GFCI receptacle(s) are not recommended since the System 83 Plus™ contains an electrostatic
isolation transformer that reduces leakage to ground and may trip the hospital breaker, the System 83
Plus'™ breaker or both.
CAUTION:
Do not connect or operate System 83 Plus™ using an emergency generator outlet because it may
create current surges that may damage computer components.
CAUTION:
The operator must perform daily checks of all in-line disc filters for all flushing adapters to the
endoscopes. Replace as required or every six (6) weeks and document the in-line disc filter change
date on the Filter and Gauge Log sheet.
CAUTION:
Tergal 800
®
CAUTION:
Only Custom Ultrasonics filters, adapters, accessories, and associated components have been
validated in the System 83 Plus™; therefore, when replacing these parts, use only OEM parts.
CAUTION:
Check HLD level in disinfectant reservoir on a per cycle basis, as prompted on the display before
performing a Wash/Disinfection cycle. Be alert to any unusual or significant loss of HLD. Be certain
to confirm HLD's MEC according to manufacturer's labeling requirements.
CAUTION:
Operator of device should refer to the HLD and Detergent handling and use requirements when
operating the device and for addressing spills. (Refer to the product material Safety Data Sheet)
CAUTION:
The System 83 Plus™ must not be connected to the facilities network.
CAUTION:
System 83 Plus™ software updates must be authorized and performed by Custom Ultrasonics.
CAUTION:
Always refer to Scope Diagrams and Scope Adaptation Charts for all endoscopes when selecting
flushing adapters for endoscope models as adaptation will differ between endoscope models.
Endoscope adaptation information can be located at: customultrasonics.com.
CAUTION:
All System 83 Plus™ maintenance must be maintained according to Custom Ultrasonics labeling
instructions to maintain the integrity and safety of the System 83 Plus™. Not following all scheduled
Custom Ultrasonics maintenance in this manual may result in damage or unsafe operating conditions
of the System 83 Plus™.
CAUTION:
Endoscopes, including duodenoscopes, should be inspected before reprocessing in the System 83
Plus™ and maintained as per the endoscope manufacturer's labeling requirements.
CAUTION:
Do not place any objects other than flexible endoscopes and endoscope accessories into the
processing chamber of the System 83 Plus™.
CAUTION:
Do not place anything on the System 83 Plus™ lids. Be certain that nothing can obstruct the lid from
closing (Example: endoscopes, valve cover, adapters etc.).
CAUTION:
Follow all facility protocols for personal protective equipment requirements PPE, such as: Eye/Mask
Protection, Rubber Gloves, and Protective Gown.
CAUTION:
Do not let the air pump remain operating while unattended.
MAN–202
is a detergent; it is NOT a High-Level Disinfectant.
Revision 03082022
System 83 Plus™ Washer/Disinfector
Service Manual
Page 7 of 54