Installation Requirements
The System 83 Plus™ should be inspected prior to installation for damage. The healthcare facility should move the
System 83 Plus™ and associated components to a secure area until installation occurs.
In-Service Training Requirements
Following delivery of a new System 83 Plus™, the medical facility must schedule a date for the training of Operating
and Bio-Medical personnel by a Custom Ultrasonics technician. During the in-service training, personnel will be
instructed on the proper operation of the System 83 Plus™ and the identification and use of endoscope adapters on
the individual flexible endoscope models in inventory at the time of the in-service. At the conclusion of this training,
all personnel in attendance are required to sign off on in-service documentation that they have been trained on how
to properly adapt each endoscope model identified in their facility inventory, as well as additional procedures and
information reviewed during the training presentation. For personnel assigned the responsibility for reprocessing
flexible endoscopes, attendance will assure all sections of the training program are reviewed and the handouts, charts
and instructions are understood and retained for reference.
A preventative maintenance and service training is provided for maintenance and/or biomedical/maintenance
personnel at the time of in-service. At the conclusion of this training, all personnel in attendance are required to sign
off on In-Service documentation that they have been trained on how to properly maintain and repair the System 83
Plus™.
WARNING:
The System 83 Plus™ must be operated by trained personnel with the necessary understanding in the
operations of the System 83 Plus™, and the requirements for the reprocessing of endoscope(s).
Service Personnel Qualifications
Custom Ultrasonics technicians, Authorized Service Providers, and Biomedical personnel that are appropriately
trained by Custom Ultrasonics are recommended to perform service on the System 83 Plus™. These personnel
receive product specific training on service and are best equipped to assure the device performs as intended.
Thoroughly review the following before use:
•
Read the Operator's Manual before use of the System 83 Plus™.
Proper handling of endoscopic equipment.
•
•
Professional health and safety standards.
•
Proper use of personal protective equipment (PPE).
Preventative Maintenance Requirements
It is imperative that proper preventative maintenance guidelines are strictly followed in accordance with Custom
Ultrasonics preventative maintenance instructions to ensure the proper operation of the System 83 Plus™. Please
contact Custom Ultrasonics for a list of Authorized Service Providers (ASP) in your area.
Warranty Requirements
Warranty period is one (1) year from the date of installation of the System 83 Plus™ or fifteen (15) months from the
date of shipment. Failure to meet all requirements could void the manufacturer warranty. Damage caused by facilities
personnel or utilities will not be covered.
MAN–202
System 83 Plus™ Washer/Disinfector
Revision 03082022
Service Manual
Page 14 of 54