Nbp/Spo2 Interlock; Asy/Vf Alarms - Dräger Infinity Vista Instructions For Use Manual

Table of Contents

Advertisement

C D
S
EFAULT
ETTINGS AND
NBP/SpO
Interlock
2
When NBP/SpO
measurement in order to avoid nuisance SpO
in the limb.
Turn NBP/SpO
1. Press the
2. Click on
3. Click on
4. Enter the Biomed password and click the knob again.
5. Click on
6. Click on
7. Turn the function

ASY/VF Alarms

Set alarms for asystole and ventricular fibrillation as either Always ON or Follow HR.
If you select Follow HR, ASY and VF alarm settings follow those settings selected for
heart rate alarms in the Alarm Limits table. If you select Always ON, the monitor
always alarms for asystole and ventricular fibrillation, even if HR alarms and
arrhythmia monitoring are turned off.
Set ASY/VF Alarms as follows:
1. Press the
2. Click on
3. Click on
4. Enter the Biomed password and click the knob again.
5. Click on
6. Click on
7. Select
C-16
B
IOMEDICAL
Interlock is ON, SpO
2
Interlock on as follows:
2
Menu
fixed key.
.
Monitor Setup
.
Biomed
.
Monitor Setup
NBP/SpO
Interlock
2
or
ON
OFF
fixed key.
Menu
.
Monitor Setup
.
Biomed
.
Monitor Setup
ASY/VF Alarms.
or
Always ON
Follow HR
Infinity Vista
S
UPPORT
alarms are disabled during an NBP
2
alarms caused by impeded circulation
2
.
.
.
VF6