Monitoring Specifications

ECG
Accessories
Input Connector
Input Parameters
Number of Channels
Sweep Speed
Resolution
Measurement Range
Measurement Accuracy
Response Time
Report Interval (nominal)
Notch Filter Frequency
Monitor ±3 dB Bandwidth
Dynamic Range AC
Dynamic Range DC
Resolution
Electrode Off Impedance
Electrode On Impedance
Calibration Spike
Defibrillation Protection
Degree of Protection Against
Electric Shock
Pacer Pulse Detection
VF6
3-,5-, and 6-Lead patient cable
2-pin connector to MultiMed pod
3-Lead (RA, RL, LA and LL electrodes)
5-Lead (RA, RL, LA, LL and CHEST electrodes)
6-Lead (RA, RL, LA, LL and 2 CHEST electrodes)
2
25 mm/s
1/min.
15 to 300 1/min.
±5 beats/minute or ±2%, whichever is greater
< 7 seconds for a step change of 80 to 120 1/min.
< 13 seconds for a step change of 80 to 40 1/min.
• Response time is measured from the onset of the first QRS
complex of the new rate to the time the measurement reads
a value that is the original rate plus 63% of the change.
2.1 seconds
50/60 Hz
0.5 Hz to 28 Hz (50 Hz)
0.5 Hz to 40 Hz (60 Hz)
±10 mV minimum
± 1 V minimum
10.4 PV
> 100 M:
< 40 M:
1 mV ±5% amplitude, 20 msec duration
In accordance with IEC 601-2-27
CF
On active ECG lead
Infinity Vista
O
VERVIEW
D-7