Oxycrg Monitoring (Neonatal Option); Displaying Ocrg Waveforms - Dräger Infinity Vista Instructions For Use Manual

Table of Contents

Advertisement

11 R
ESPIRATION

OxyCRG Monitoring (Neonatal Option)

The monitor offers a neonatal Oxycardiorespirogram (OCRG or OxyCRG) provided
the option is enabled on your monitor (for more information, see your Dräger
representative). Neonatal OxyCRG is a non-invasive procedure to display a beat-to-
beat heart rate (bbHR) trend, an SpO
simultaneously on a dynamically updated display. (OCRG monitoring requires that
sensors and electrodes for SpO

Displaying OCRG Waveforms

The OxyCRG waveforms are displayed in the second and third waveform channels.
The second waveform channel shows the beat-to-beat Heart Rate (bbHR) trend and
the SpO
trend waveforms. The bbHR trend has a range of 65 to 290. The SpO
2
has a range from 70% to 100%. The third waveform channel shows a time-
compressed respiration waveform.
STEPS: Displaying the OCRG Waveforms
1. Make sure the Neonatal monitoring mode is selected.
2. Click on the
3. Click on
4. Select
11-12
M
ONITORING
, HR, and respiration are attached to the patient.)
2
or
second
third waveform
.
Waveform
and click the knob.
OCRG
Infinity Vista
trend, and a respiration waveform
2
channel.
trend
2
VF6

Advertisement

Table of Contents
loading

Table of Contents