Arrhythmia Setup
The Arrhythmia Setup table lets you set Rate and Count limits for certain arrhythmia
parameters and turn alarms and alarm recordings on or off. Call up the Arrhythmia
Setup table as follows:
1. Click on the
2. Click on
When you turn full arrhythmia monitoring on, the Arrhythmia Setup table spans 2
pages. Click on one of the arrow keys in the upper left-hand corner to view additional
parameters.
NOTE:
z
Alarm settings for PVC can be selected in the Alarm Limits table (see the
chapter Alarms and Messages).
z
By default, alarms for asystole and ventricular fibrillation are always on;
however, these alarms can be disabled in the password-protected Biomed menu
(see the appendix Default Settings and Biomedical Support).
WA R N I N G : I f t h e m o n i t o r d i s p l a y s t h e b a n n e r H R , A S Y, V F
O ff , t h e r e i s n o a l a r m a n n u n c i a t i o n f o r a s y s t o l e a n d
v e n t r i c u l a r f i b r i l l a t i o n .
VF6
parameter box.
HR
.
Arrhythmia Setup
Infinity Vista
A
S
RRHYTHMIA
ETUP
9-5