Respiration (continued)
Resolution
Measurement Range
Measurement Accuracy
Response Time
Report Interval
Total resistance
Input Dynamic Range
Bandwidth
Impedance Measuring Current
Pulse Oximetry (SpO
2
Parameter Display
Measuring Method
Measuring Range
Calibration Range
Display Range
Display Update Period
Maximum Hold from Previous
Update
VF6
1 breath per minute
0 to 155 breaths per minute
± 1 breath per minute, or ± 2%, whichever is greater
After a step change in respiration rate, it takes 36 seconds to
stabilize to a new respiration rate of 10 breaths per minute
and 18 seconds to stabilize to a new respiration rate of 20
breaths per minute.
2 seconds
0 to 3600 :
AC: ± 17.3 : ± 10%
DC: 0 : to 3.6 k: (internal check for high impedance)
0.25 Hz to 3.5 Hz
65 PA RMS ±10% nominal at 48kHz ±1%, rectangular wave
)
Saturation (%SpO
Absorption-spectrophotometry
SpO
:
2
Pulse rate:
70-100%
0-100%
2 seconds
30 seconds (in the event of artifact or other error)
Infinity Vista
), pulse rate
2
1 - 100%
30 - 250 1/min
O
VERVIEW
D-9