Patient Category - Dräger Infinity Vista Instructions For Use Manual

Table of Contents

Advertisement

4 A
/D
DMISSION
ISCHARGE

Patient Category

NOTE: The currently selected patient category is indicated between the first and
second waveform channels next to the parameter boxes.
STEPS: Selecting the Patient Category
1. Call up the Patient Admit menu (Menu > Admit/Discharge > Patient Admit,
see above).
2. Click on
3. Dial in the desired patient category (Adult, Pediatric, Neonate) and click the
knob.
When you change the patient category, the monitor:
z
Clears all current alarms, including their messages.
z
Stops any NBP measurement in progress.
z
Deletes stored trend data, events, and recordings.
z
Returns alarm limits to their category-specific default settings (see appendix
Default Settings and Biomedical Support).
z
Returns waveform display scales (sizes) to their category-specific default
settings (see appendix Default Settings and Biomedical Support).
z
Sets the display range for trend values according to the selected patient
category (see Trends chapter).
4-4
/T
RANSFER
.

Patient Category

Infinity Vista
VF6

Advertisement

Table of Contents
loading

Table of Contents