Electrosurgery and Defibrillation Safety

The monitor is protected against high-frequency interference from electrosurgery
units (ESU) and discharges from defibrillators, as well as against 50- and 60-Hertz
power line interference.
WA R N I N G : T h e m o n i t o r i s n o t p r o t e c t e d a g a i n s t h i g h -
f r e q u e n c y i n t e r f e r e n c e f r o m d i a t h e r m y e q u i p m e n t .
CAUTION: During Electrosurgery, observe the following guidelines to minimize ESU
interference and provide maximum user and patient safety:
z
Use the ESU block to connect ECG cables.
z
Keep all transducers and intermediate cables off of earth ground and away
from the ESU knife and return wires.
z
Use the SpO
z
Use rectal temperature probe sheaths to cover any internally placed
temperature sensors.
z
Always use the accessories designed for ESU environments.
z
If pacer detection is on, the ESU interference may be detected as pacer spikes
displayed on the ECG.

Pacemaker Safety

WA R N I N G : R a t e m e t e r s m a y c o n t i n u e t o c o u n t t h e pa c e m a k e r
r a t e d u r i n g o c c u r r e n c e s o f c a r d i a c a r r e s t o r s o m e
a r r h y t h m ia s . D o n o t r e l y e n t i r e l y u p o n r a t e m e t e r a l a r m s .
K e e p pa c e m a k e r pa t i e n ts u n d e r c l o s e s u r v e i l l a n c e .
WA R N I N G : M a k e s u r e t h a t pa c e r d e t e c t i o n i s t u r n e d o f f f o r
pat i e n ts w i t h o u t pa c e m a k e r s , a n d t u r n e d o n f o r pat i e n ts w i t h
pa c e m a k e r s .
1-6
pulse rate instead of the ECG to determine the heart rate.