ECORDINGS

Recording Waveforms

To print real-time monitoring data, two types of recordings are available:
z

Timed Recordings.

z
Continuous Recordings.
For both recording types, the recording speed is 25 mm/s.
Timed and continuous recordings include up to two waveforms. If one of the
waveforms is hidden by a menu, you can still request a recording. If a cascaded
waveform is displayed, only the waveform from the first channel prints out.
NOTE: Timed and continuous recordings cannot be started while the Alarm Limits
table, trends or diagnostic logs are displayed. To start recording, exit those displays
before requesting a recording.

Timed Recordings

Timed Recordings print a total of 20 seconds of patient data, of which about 10
seconds occurred prior to the recording request (Delay), and about 10 seconds after
the recording request.
If no recorder/printer is available, the monitor stores the recording (see Stored
Recordings, below). You can cancel the storage within 5 seconds of the initial
recording request.
The monitor cancels a timed recording if you modify the display, size or scale of a
waveform while the recording is ongoing.
STEPS: Starting a Timed Recording
1. Press the monitor's
2. To cancel the recording, press the
recorder while printing is in progress.