Data Transfer

You can transfer patient data between monitors in the following ways:
z
By sending data to a different monitor over the Infinity network.
z
By copying data onto a memory card and then loading it into a different
monitor (stand-alone option).
The transfer of patient data involves a source monitor and a destination monitor. The
source monitor is the unit from which the data is transferred. The destination monitor
is the unit receiving the data. Transferred patient data includes the patient's name, ID,
admit date, patient category, as well as stored trends.
NOTE:
z
When a data transfer takes place between two monitors whose selected units of
measure do not match (i.e.°C/°F, mmHg/kPa, mm/mV), the destination monitor
adopts the units of measure that were selected at the source monitor.
z
When a data transfer takes place between two monitors with different enabled
software options, the destination monitor accepts data only for those options
which are enabled at the destination monitor (i.e. IBP), and ignores the rest.
z
For data transfer between Infinity Vista monitors, lead V+ is transferred only if
both the source and the destination monitors are using a 6-lead ECG cable. Use
of a smaller lead set cable only transfers data corresponding to the smaller lead
set. The remaining data is lost.
z
Stored events cannot be transferred.

Transfer Across the Network

NOTE: Before data transfer can take place over the network, the source monitor must
be placed into Standby mode.
STEPS: Transfer Across the Network
At the source monitor:
1. Place the source monitor into Standby.
At the destination monitor:
1. Press the
2. Click on
3. Click on
4-8
/T
RANSFER
fixed key.
Menu
.
Admit/Discharge
. The Transfer menu opens.
Transfer