Saving a Patient Setup

You can save and re-use the current monitoring configuration. This saved
configuration is automatically restored when a new patient of the same category (e.g.,
pediatric) is admitted.
The monitor saves the following settings: Waveform channel assignments and scales,
alarm limits and on/off status, IBP pressure labels, arrhythmia monitoring settings
(adult or pediatric mode only), respiration mode and markers, apnea time, coincidence
alarm and pacemaker setting and detection.
STEPS: Saving a Patient's Setup
1. Press the
2. Click on
3. Click on
4. Dial in the Biomed password and click the knob.
5. Click on
6. Click on
The monitor saves the current monitoring setup for future use. The setup is available
even after the current patient has been discharged from this monitor.
NOTE: You cannot save a setup when an OCRG is displayed on the screen. Exit the
OCRG display before saving.

Locked Options and Demo Mode

The Locked Options menu gives you access to the Demo mode and to a number of
monitoring options that are available for purchase. These options and their order
numbers are listed in the appendix Options and Accessories.
Using the Demo Mode
The monitor provides a demonstration mode to allow the presentation of features and
functions during a product introduction. Before demonstrating or testing the monitor,
disconnect all patient cables from the monitor.
NOTE: In Demo mode, the All Alarms Off key is without function and creates an
error tone when pressed.
VF6
fixed key.
Menu
.
Monitor Setup
.
Biomed
.
Save Setups
, select
Save Setup
Confirm
Infinity Vista
B
IOMEDICAL
(or
) and click the knob.
Cancel
S
UPPORT
C-11