Required Test Equipment; Test Setup - Agilent Technologies NFA Series Performance Verification And Calibration Manual

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Table 2-17
Equipment description
Synthesized Sweeper
Adapter 1
Adapter 2
RF Cable
Figure 2-16
NOISE FIGURE
ANALYZER

Required Test Equipment

Equipment required for Gain Measurement Uncertainty test
Critical specification for equipment
substitution
Frequency Range: 50 MHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -46 dBm
N8973A N Type (M) to 3.5mm (F)
N8974/5A 3.5mm Precision (F) to 3.5mm (F)
83620/30 3.5mm Precision (F) to 3.5mm (F)
83640/50 2.4mm Precision (F) to 3.5mm (F)
50 MHz 3.5mm (M) to 3.5mm (M), 61cm (24
in.)

Test Setup

Gain Measurement Uncertainty Test Setup
Input
Type N Cable
Chapter 2
Gain Measurement Uncertainty
Recommended model
83620/30/40/50B Option
001 and 008
1250-1744
1250-1749 or 83059B
1250-1749 or 83059B
11901B
11500E
SYNTHESIZED
SWEEPER
RF Output
Adapter
Test Descriptions
65

Advertisement

Table of Contents
loading

Table of Contents