Agilent Technologies NFA Series Performance Verification And Calibration Manual page 66

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Test Descriptions
Noise Source Supply Accuracy
Step 8. Set the noise source on (press the
Noise Source On
Press
then
Enter
Step 9. Enter the digital multimeter reading in the Test Record as the Noise Source Supply
On measurement.
Step 10. Press the
Preset
Step 11. Verify that the measured values are within the published specifications.
54
System
). Press
Enter Password
and
Service
Noise Source On
key to return to the default Noise Source Supply State.
key, then
,
and
More
More
then enter the service password -2010.
).
. Press
Service
Chapter 2