Agilent Technologies NFA Series Performance Verification And Calibration Manual page 42

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Step 9. Select the relevant calibration kit in use from the Cal menu, as the calibration kit
NOTE
Step 10. Select an S11 port calibration (press the
Step 11. Connect the Open to the test port or, if fitted, the adapter and perform the open
Step 12. Connect the Short to the test port or, if fitted, the adapter and perform the open
Step 13. Connect the Load to the test port or, if fitted, the adapter and perform the open
Step 14. Save the calibration (press the
Test Descriptions
Input VSWR
selection is dependant on the NFA being tested (press the
selection and
).
Return
The connector type applies to the test port or adaptor if fitted and not the NFA under
test.
).
Reflection 1-port
calibration (press
Opens
calibration (press
Shorts
calibration (press
Load
30
key, then
Cal
and then select the relevant menu key).
and then select the relevant menu key).
, select the relevant menu key
Save/recall
key and then
key,
Cal
Calibration Kit
and
Calibrate Menu
and then done 1-port CAL
Save State
).
Chapter 2
).