Agilent Technologies NFA Series Performance Verification And Calibration Manual page 74

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Table 2-14
Table 2-15
Table 2-16
Test Descriptions
Noise Figure Range and Accuracy
4.5-6.5 dB Noise Source ENR Test Record
Actual NF range
Measured NF range
Instrumentation Uncertainty:
14-16 dB Noise Source ENR Test Record
Actual NF range
Measured NF range
Instrumentation Uncertainty:
Range
0.125 to 11
dB
Range
12 to 15
dB
20-22 dB Noise Source ENR Test Record
Actual NF range
Measured NF range
Instrumentation Uncertainty:
Range
0.125 to 11
dB
Range
12 to 22
dB
62
=
5 dB ENR - 10LOG (10^ (Actual
Attenuation
/10) -1)
dB
=
5 dB ENR - 10LOG (10^ (Measured
Attenuation
/10) -1)
dB
=
Actual NF - Measured NF
=
15 dB ENR - 10LOG (10^ (Actual
Attenuation
/10) -1)
dB
=
15 dB ENR - 10LOG (10^ (Measured
Attenuation
/10) -1)
dB
=
Actual NF - Measured NF
=
Actual NF - Measured NF +
Instrumentation Uncertainty at 11 dB
(re-reference)
=
22 dB ENR - 10LOG (10^ (Actual
Attenuation
/10) -1)
dB
=
22 dB ENR - 10LOG (10^ (Measured
Attenuation
/10) -1)
dB
=
Actual NF - Measured NF
=
Actual NF - Measured NF +
Instrumentation Uncertainty at 11 dB
(re-reference)