Test Setup; Test Procedure - Agilent Technologies NFA Series Performance Verification And Calibration Manual

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Figure 2-7
NOISE FIGURE
ANALYZER
Adapter 1
NOTE
Step 1. Connect the equipment as shown in Figure 2-7 on page 40.
Step 2. Set the Synthesized Sweeper as follows:
Test Descriptions
Frequency Accuracy

Test Setup

Frequency Accuracy test setup
Ref In
Ref In
Input
RF Cable

Test Procedure

Throughout the Frequency Accuracy Test Procedure the term 'the Worksheet' refers
to the Frequency Accuracy Test Worksheet on page 45 and the term 'the Test
Record' refers to the Frequency Accuracy Test Procedure Test Record detailed in
the relevant Appendix for the model number being tested.
1. Press the
Preset
key and wait for the preset routine to complete.
2. Set the power level to -70dBm (press the
3. Switch
RF ON
40
BNC Cable
SYNTHESIZED
SWEEPER
10 MHz
RF Output
Adapter 2
Power Level
FREQUENCY
STANDARD
key and
,
,
,
-
7
0
dB(m)
Chapter 2
).

Advertisement

Table of Contents
loading

Table of Contents