Table 1-4
Test no.
Performance Verification Test
1
10 MHz Out Frequency Reference Accuracy
2
10 MHz Out Frequency Reference Adjustment
3
Input VSWR
4
Frequency Accuracy
5
Noise Source Supply Accuracy
6
Noise Figure Range and Accuracy
7
Gain Measurement Uncertainty
8
Instrument Noise Figure
9
Measurement Jitter
Preparing for Calibration and Performance Verification Test

Performance Verification and Adjustment Procedures

Performance Verification and Adjustment Procedures
To perform a calibration run, the performance verification tests listed in Table 1-4
below must be completed.
If any of the performance verification tests fail, perform the corresponding
calibration adjustment listed in Table 1-4.
If the performance verification tests fail and no calibration adjustment procedure is
available, the instrument must be returned to Agilent Technologies for repair. Please
contact your local Agilent Technologies Customer Sales and Service Office for
replacement parts or repair service information. See the NFA User's Guide for the
address of your nearest office.
Verification and adjustment listings
Chapter 1
Calibration Adjustments