Agilent Technologies NFA Series Performance Verification And Calibration Manual page 149

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Instrument Noise Figure Test Record
Frequency (MHz)
10.00 to 26500
10.00
30.00
60.00
100.00
300.00
600.00
1000.00
1500.00
2000.00
3000.00
4000.00
5000.00
6000.00
7000.00
8000.00
9000.00
10000.00
11000.00
12000.00
Appendix C
Instrument Noise
Figure (dB)
Overall Frequency
Range Result
Model N8975A: Test Records
Instrument Noise Figure Test Record
Specification (dB)
Test Line Limit
< 4.400dB
< 4.440dB
< 4.470dB
< 4.520dB
< 4.760dB
< 5.110dB
< 5.570dB
< 6.120dB
< 6.740dB
< 7.910dB
< 10.50dB
< 10.50dB
< 10.50dB
< 10.50dB
< 10.50dB
< 10.50dB
< 10.50dB
< 10.50dB
< 10.50dB
Pass/Fail
139

Advertisement

Table of Contents
loading