Step 12. Perform the frequency accuracy calculations either using a spreadsheet such as
Step 13. Repeat steps 4 to 12 above for the remaining frequencies, spans and bandwidths as
Step 14. Reset the IF Gain (press the
NOTE
Test Descriptions
Frequency Accuracy
®
Microsoft
Excel (see page 49), or manually (see page 50).
listed in the data into the Test Record. Ensure the measured values are within its
published specification.
service password
-, 2, 0,1 ,0
Gain (Off)
.
The DSP bandwidths are verified at the narrowest bandwidth only. This is due the
fact that they use digital sample processing (DSP), therefore it is only necessary to
verify the narrowest bandwidth.
44
key, then
System
More
press
,
and then press
Enter
Service
,
, and
. Enter the
More
Service
and
IF Test
Chapter 2
Fixed IF