Figure 2-2
NOISE FIGURE
ANALYZER
NOTE
NOTE

Test Setup

Before starting the 10 MHz Out Frequency Reference Adjustment, connect the
equipment as shown in Figure 2-2.
10 MHz Out Frequency Reference Adjustment Test Setup
10 MHz
Ref Out
UNIVERSAL
COUNTER
BNC Cable

Test Procedure

Before performing any adjustments ensure that the 10 MHz Frequency Reference
Accuracy test has been done and that the measured frequency is outside the
published specification.
The NFA must be on and in internal frequency mode for at least 1 hour before you
start the test.
Chapter 2
10 MHz Out Frequency Reference Adjustment
BNC Cable
10 MHz
Ref In
Channel 1
Test Descriptions

FREQUENCY

STANDARD