Required Test Equipment - Agilent Technologies NFA Series Performance Verification And Calibration Manual

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Table 2-11
Equipment description
Synthesized Sweeper
Attenuator/Switch
Driver
1 dB Step Attenuator
Adapter 1
Adapter
RF Cable (X 2)
Test Descriptions
Noise Figure Range and Accuracy

Required Test Equipment

Noise Figure Range and Accuracy test equipment
Critical specification for equipment
substitution
Frequency Range: 50 MHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -54dBm to -65dBm
Compatible with 8494G/H Programmable step
attenuators
Attenuation Range: 0 to 11dB
Frequency Range: 50 MHz
Connectors: 3.5mm (F)
Calibrated at 50 MHz with accuracy of 0.010 dB
N8973A N Type (M) to 3.5mm (F)
N8974/5A 3.5mm Precision (F) to 3.5mm (F)
83620/30 3.5mm Precision (F) to 3.5mm (F)
83640/50 2.4mm Precision (F) to 3.5mm (F)
50 MHz 3.5mm (M) to 3.5mm (M), 61cm (24 in.)
56
Recommended model
83620/30/40/50B
Option 001 and 008
11713A
8494G/H Option 002
1250-1744
1250-1749 or 83059B
1250-1749 or 83059B
11901B
11500E
Chapter 2

Advertisement

Table of Contents
loading