Figure 2-13
NOTE
Step 1. Connect the equipment as shown in Figure 2-13.
Step 2. Select the DC range on the digital multimeter (press
Step 3. Ensure that the preset is set to factory settings (press the
Step 4. On the NFA press the
Step 5. Set sweep to single (press
Step 6. Set the noise source off (press the
Step 7. Enter the digital multimeter reading into the Test Record as the Noise Source Supply

Test Setup

Equipment setup for Noise Source Supply Accuracy test
DIGITAL
MULTIMETER
Adapter on
(2 Wire) HI/LO

Test Procedure

Throughout the Noise Source Supply Accuracy Test Procedure the term 'the Test
Record' refers to the Noise Source Supply Accuracy Test Record detailed in the
relevant Appendix for the model number being tested.
menu).
and
Power On (Preset)
Preset
Enter Password
then enter the service password -2010. Press
and
).
Noise Source Off
Off measurement.
Chapter 2
Noise Source Supply Accuracy
+28V Noise
Source Supply
BNC Cable
).
Preset (Factory)
key and wait for the routine to finish.
,
Sweep
Sweep Mode: single
key, then
System
Test Descriptions
NOISE
FIGURE
ANALYZER
in the
DCV
FUNCTION/RANGE
System
key, then
).
,
and
More
More
Service
Enter
then
More
,
. Press
Service
53