Test Setup; Test Procedure - Agilent Technologies NFA Series Performance Verification And Calibration Manual

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Figure 2-14
NOISE FIGURE
ANALYZER
Adapter 1
NOTE
Step 1. Connect the equipment as shown in Figure 2-14.
Step 2. Set the Synthesized Sweeper as follows:
Step 3. Set the Switch Driver as follows:

Test Setup

Equipment required for Noise Figure Range and Accuracy test
ATTENUATOR
SWITCH DRIVER
Input
PRECISION 1dB
STEP ATTENUATOR
RF Cable

Test Procedure

Throughout the Noise Figure Range and Accuracy Test Procedure the term 'the
Worksheet' refers to the Noise Figure Range and Accuracy Test Worksheet on page
60 and the term 'the Test Record' refers to the Noise Figure Range and Accuracy
Test Record detailed in the relevant Appendix for the model number being tested.
1. Press the
key.
Preset
2. Set the frequency to 50 MHz (press the
3. Set the power level to -54 dBm (press the
4. Switch
RF ON
Chapter 2
Test Descriptions
Noise Figure Range and Accuracy
SYNTHESIZED
SWEEPER
Control
Adapter 2
Cable
Attenuator X
RF Cable
CW
key and
5
,
0
,
MHz
Power Level
key and
.
-
,
5
,
4
,
dB(m)
.
57