Model N8973A: Test Records
Instrument Noise Figure Test Record
Instrument Noise Figure Test Record
Frequency (MHz)
10.00 to 3000
10.00
30.00
60.00
100.00
300.00
600.00
1000.00
1500.00
2000.00
3000.00
114
Instrument Noise
Figure (dB)
Overall Frequency
Range Result
Specification (dB)
Test Line Limit
< 4.400dB
< 4.440dB
< 4.470dB
< 4.520dB
< 4.760dB
< 5.110dB
< 5.570dB
< 6.120dB
< 6.740dB
< 7.910dB
Pass/Fail