Table 2-6
Equipment description
Vector Network Analyzer 1
Vector Network Analyzer 2
N Type Calibration Kit
3.5mm Calibration Kit
Test Descriptions
Input VSWR

VSWR Test Specification

10 MHz to 500 MHz
> 500 MHz to 1500 MHz
> 1500 MHz to 3000 MHz
> 3000 MHz to 6700 MHz
> 6700 MHz to 20000 MHz
> 20000 MHz to 26500 MHz

Related Adjustments

None

Required Test Equipment

Test Equipment required for Input VSWR test
Critical specification for
equipment substitution
10 MHz to 3 GHz
3 GHz to 26.5 GHz
Impedance: 50
Frequency Band: 10 MHz to
c
3 GHz
UUT: N8973A
Impedance: 50
Frequency Band: 10 MHz to
3 GHz
UUT: N8974A and N8975A
26
N8973A
N8974A
< 1.5:1
< 1.5:1
< 1.7:1
< 1.7:1
< 1.8:1
< 1.8:1
N/A
< 1.3:1
N/A
N/A
N/A
N/A
Recommended model
8753ES or 8753ET, Option 004
8722ES or 8722ET, Option 004
Ω
85032B
Ω
85033D
N8975A
< 1.5:1
< 1.7:1
< 1.8:1
< 1.3:1
< 2.1:1
< 2.4:1
a