Step 1. Open Excel.
Step 2. Open the .csv file from the A: drive. From the
Figure 2-10
Step 3. Increase the column widths to 20 to view the data more accurately.
Step 4. Enter the following formula into cell D1 = SUM(B$1:B1), copy this formula and
Step 5. Enter the following formula into cell D202 =D201/2. This value returns the ½
Performing the frequency accuracy calculations using the CSV file in a
spreadsheet
The following procedure uses Microsoft Excel.
(A:). Select files of
Type:
Example Excel File
Column A contains the measured frequency. Column B contains the corresponding
Linear P
value. Disregard Column C null terminator information.
hot
paste it from cell D2:D201.
Power reading.
Chapter 2
File
Text Files (*.prn; *.txt; *.csv).
Test Descriptions
Frequency Accuracy
menu select
and
Open
Look in:
49