The microprocessor-controlled hydraulic unit offers the following advantages
Approximation of the physiological gait pattern
Stability while standing and walking
Adaptation of product characteristics to various surfaces, inclines, gait situations and walking speeds
Essential performance of the product
Stability in the stance phase

2.3 Combination possibilities

This product can be combined with the following Ottobock components:
Prosthetic hip joints
Modular prosthetic hip joint: 7E7
Monocentric prosthetic hip joint: 7E9
Adapters
4R104=60 double adapter, sliding
4R104=75 double adapter, sliding
Rotation adapter: 4R57, 4R57=*
4R41 lamination anchor with pyramid receiver
4R111 lamination anchor with pyramid receiver
4R89 lamination anchor with pyramid adapter
Tube adapter
2R57 tube adapter
2R67 tube adapter with torsion unit
Cosmetic cover/protector
Foam cover: 3S26
3F1=1 Functional cosmesis C-Leg
99B120=* Functional stocking
Prosthetic feet
The maximum allowable patient weight depends on the foot size.
1D10 Dynamic foot
1D11 Dynamic foot (women)
1M10 Adjust
1C10 Terion
1C11 Terion K2
1A30 Greissinger plus
1C30 Trias
1C30-1 Trias
1D35 Dynamic Motion
1C40 C-Walk
1C50 Taleo
1C51 Taleo Vertical Shock
1C52 Taleo Harmony
1C53 Taleo Low Profile
1C58 Taleo Side Flex
1
Note the Ottobock system height
INFORMATION
Calculating the Ottobock system height for the F22, VS2, FS5, RS2-00, RS2-K1 prosthetic feet
To calculate the Ottobock system height (e.g. for input in the adjustment software), the build height according to
the technical data for the listed prosthetic feet has to be reduced by about 18 mm.
Example: The build height of the "F22" prosthetic foot in size 26 is 117 mm.
Therefore, the system height is: 117 mm – 18 mm = 99 mm. This is only a reference value. Therefore, measure
and verify the distances on the patient before shortening the tube adapter.
C-Leg 3C98-3/3C88-3
3D
Helix
prosthetic hip joint: 7E10
4R116 lamination anchor with pyramid adapter
4R43 lamination anchor with threaded connector
4R111=N lamination anchor with threaded connect­
or
4R40 torsion adapter
4R118 adapter plate
4X860=* C-Leg Protective Cover (w/o shield)
4P862 C-Leg guard
4P863* Shield Insert
1C60 Triton
1C61 Triton Vertical Shock
1C62 Triton Harmony
1C63 Triton Low Profile
1C64 Triton Heavy Duty
1C68 Triton side flex
1E56 Axtion
1E57 Lo Rider
1B1 Meridium
1B1-2 Meridium
1A1-2 Empower
F22 Maverick Comfort AT
1
VS2 Promenade
1
FS5 Thrive
1
RS2-00 Runway
, RS2-K1 Runway HX
Product description
1
1