3 Application

3.1 Indications for use

The product is to be used solely for lower limb prosthetic fittings.

3.2 Area of Application

CAUTION
Excessive strain on the product
Falling due to breakage of load-bearing components
Use the product exclusively according to its MOBIS classification (see Fig. 
Use the product exclusively for the approved amputation levels 
5
(see Fig. 
on page 6).
The patient has the ability or the potential to use the prosthesis for the purpose of moving or proceeding
slowly on level floors. The amount of time and the distance he/she can walk are seriously limited due to
his/her condition.
The patient has the ability or the potential to walk slowly with the prosthesis and to negotiate low environ­
mental obstacles like curbs, single steps or uneven surfaces. The amount of time and the distance he/she
can walk are seriously limited due to his/her condition.

3.3 Combination Possibilities

CAUTION
Unallowable combination of prosthetic components
Fall due to breakage or deformation of the product
Note that the illustrated combinations with the pictogram 
pictogram 
are not permitted (see Fig. 
Only combine the product with prosthetic components that are approved for the patient
based on the combination table according to MOBIS (see Fig. 
2
hip joints,
 proximal adapters,
6
feet,
 weight class: see instructions for use).
Based on the instructions for use of the prosthetic components, verify that they may be com­
bined with each other.
For additional combination possibilities, please see the catalogue 646K2* or consult the manufac­
turer.

3.4 Environmental Conditions

CAUTION
Use under unallowable environmental conditions
Fall due to damaged product
Do not expose the product to unallowable environmental conditions (see the table "Unallow­
able environmental conditions" in this section).
If the product has been exposed to unallowable environmental conditions, check it for dam­
age.
If damage is apparent or in case of doubt, do not continue using the product.
Take suitable measures if required (e. g. cleaning, repair, replacement, inspection by the
manufacturer or a specialist workshop etc.).
42 | Ottobock
MG 1: indoor walker
MG 2: Restricted outdoor walker
6
on page 6).
3
 prosthetic knee joints,
4
on page 5).
1
and amputation sides 
are permitted and those with the
7
on page 7:
4
 distal adapters,
2
1
 prosthetic
5
 prosthetic

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

3r120 kd3r120 hdPuro spring 3r120Puro spring 3r120-kdPuro spring 3r120-hd

Table of Contents