Otto Bock C-Leg 3C98-3 Instructions For Use Manual page 53

Hide thumbs Also See for C-Leg 3C98-3:
Table of Contents

Advertisement

Beep signal
4 x short at intervals of
approx. 20 sec. (continu­
ously)
17.2.3 Error messages while establishing a connection with the cockpit app
Error message
Component was connec­
ted
to
another
device.
Establish connection?
Mode change failed
17.2.4 Status signals
Battery charger is connected
LED on
LED on
power sup­
battery
ply
charger
Battery charger disconnected
Beep sig­
Vibration
nal
signal
1 x short
1 x short
3 x short
Battery charge level
Battery
charger
Battery is being charged, battery charge level is less than 50%
Battery is being charged, battery charge level is over 50%
C-Leg 3C98-3/3C88-3
Error
Charging the battery outside the allowable
temperature range
Cause
The component was connec­
ted to another mobile device
An attempt was made to
switch to a different MyMode
while the component was in
motion (e.g. while walking)
A current connection to the
prosthesis was interrupted
Event
Power supply and battery charger operational
Result
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery charger. If the
beep signal sounds again, product maintenance should be carried out by an autho­
rised Ottobock Service Center.
The product can be used without restrictions. However, vibration signals may not be
generated.
Conduct the self-test again by connecting/disconnecting the battery charger. If no
beep/vibration signal is emitted after connecting/disconnecting the battery charger
again, the product must be inspected by an authorised Ottobock Service Center.
Resolution
Check whether the specified ambient con­
ditions for charging the battery are met
(see page 48).
Correction
To disconnect the original connection, tap the "OK"
button.
If the original connection is not to be disconnected,
tap the "Cancel" button.
For safety reasons, switching MyModes is only per­
mitted when components are at rest, e.g. while
standing or sitting.
Check the following points:
Distance from the prosthesis to the mobile
device
Charge level of the prosthesis
Bluetooth on the prosthesis switched on? (see
page 41)
Hold the component with the sole of the foot
facing up to make the component "visible" for
2 minutes.
Prosthesis switched on? (see page 41)
If multiple prostheses were stored, was the cor­
rect prosthesis selected?
Appendices
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

C-leg 3c88-3

Table of Contents